Tektronix AWG710B User Manual

Page of 580
Appendix B: Performance Verification (AWG710)
AWG710&AWG710B Arbitrary Waveform Generator User Manual
B-85
Pulse Response Tests (Normal Out), (except option 02)
This procedure checks the pulse response characteristics of the AWG710 Arbitrary 
Waveform Generator output waveforms at amplitudes of 1 V. 
Do the following steps to install the test hookup and set the test equipment controls:
1. Use the 50 
Ω
 SMA coaxial cable to connect the AWG710 Arbitrary Waveform 
Generator CH1 output connector to the oscilloscope CH1 input connector (see 
Figure B-48)
Figure B-48: Pulse response initial test hookup
2. Set the oscilloscope controls as follows:
3. Push UTILITY (front–panel)!System (bottom)!Factory Reset (side) 
!OK (side).
Equipment 
required
A 50 
Ω
 SMA coaxial cable and an oscilloscope (TDS820).
Prerequisites
The AWG710 Arbitrary Waveform Generator must meet the prerequisites 
listed on page B-61.
VerticalCH1
CH1 coupling. . . . . . . . . . . . . . . . . . . . . . . . . DC if applicable
CH1 scale . . . . . . . . . . . . . . . . . . . . . . . . . . . 200 mV/div
Horizontal
Sweep . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 500 ps/div
Trigger
Source. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . CH1
Slope . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Positive
Level . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0 V
Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Auto
AWG710 Arbitrary Waveform Generator
Oscilloscope (TDS820)
50
Ω
 SMA coaxial cable