Appendix B: Performance Verification (AWG710)
AWG710&AWG710B Arbitrary Waveform Generator User Manual
B-91
Event Input and Enhanced Mode Tests
These procedures check the event input signals and enhanced mode operation. 
Check Event Input with
Strobe Off
Do the following steps to install the test hookup and set the test equipment controls:
1. Use a 50
Ω
 SMA coaxial cable and a SMA(Fe)–BNC(Fe) adapter to connect 
the AWG710 Arbitrary Waveform Generator CH1 output connector to the 
oscilloscope CH1 input connector (see Figure B-54).
Figure B-54: Event input and enhanced mode initial test hookup
2. Connect the ground closure to the EVENT IN connector on the AWG710 
Arbitrary Waveform Generator rear panel.
NOTE. 
The event input check with strobe off and the strobe input check are 
structured as a continuous test. After Check Event Input with Strobe Off, the next 
test uses the connections and oscilloscope settings from the previous test.
Equipment 
required
A 50
 Ω
 SMA coaxial cable, a SMA(Fe)–BNC(Ma) adapter an oscilloscope 
(TDS700), and custom–made ground closure. See Figure B-37 for the 
connections.
Prerequisites
The AWG710 Arbitrary Waveform Generator must meet the prerequisites 
listed on page B-61.
AWG710 Arbitrary Waveform Generator
Oscilloscope (TDS700)
50 
Ω
 SMA coaxial cable
SMA(Female)-BNC
(Male) adapter