Bunn 38791.0000K User Manual

Page of 26
20
 
LCD DISPLAY 
 CALIBRATION MODE 
 DESCRIPTION
CALIBRATING THE DISPENSER
Left (Regular) Pump Cal.
Set PRIME-NORMAL-RINSE switch 
to PRIME position. Press and hold 
Regular Dispense Switch until pump 
stops (20 sec.) Refer to Field Calibration 
of the Concentrate Pumps.
NOTE: The LCA-2 LP is calibrated at the factory and does not normally need to be re-calibrated. To enter the 
Calibration Mode, Set the RUN/NIGHT/PROGRAM Switch to the PROGRAM position and then hold the “MENU” 
switch down for 10 seconds.
Left (Regular) Pump Cal.
Enter mL collected from the 20 sec. 
test above.
Default = 22
Range = 18 - 26mL
Right (Decaf) Pump Cal.
Set PRIME-NORMAL-RINSE switch to 
PRIME position. Press and hold Decaf 
Dispense Switch until pump stops (20 
sec.) Refer to Field Calibration of the 
Concentrate Pumps.
Right (Decaf) Pump Cal.
Enter mL collected from the 20 sec. 
test above.
Default = 22
Range = 18 - 26mL
Left (Regular) Water Rate
Set PRIME-NORMAL-RINSE switch to 
RINSE position. Press and hold Regular 
Dispense Switch until flow stops (20 
sec.) Refer to Field Calibration of the 
Concentrate Pumps.
Left (Regular) Water Rate
Enter Liters collected from the 20 sec. 
test above.
Default = 1.08
Range = 0.95 - 1.12 Liters
38791 072307
MENU
(-)
(+)
C 1 r u n
MENU
(-)
(+)
C 1
2 2
MENU
(-)
(+)
C 2 r u n
MENU
(-)
(+)
C 2
2 2
MENU
(-)
(+)
C 3 r u n
MENU
(-)
(+)
C 3 1.0 8