Sierra 840 User Manual

Page of 74
Series 830/840/860 Instruction Manual 
Chapter 4 Maintenance
 
IM-83/84/86-H 
4-1
 
Chapter 4  Maintenance 
Side-Trak
 and Auto-Trak
 transducers essentially require no 
scheduled maintenance other than periodic flow path cleaning if 
the gas is dirty. If an in-line filter is used, the filtering element 
should be periodically be replaced or ultrasonically cleaned. In-
cluded in this chapter are general instructions for: 
 
 
•  Flow Path Maintenance ...............page 4-1 
 
•  Sensor Maintenance.....................page 4-6 
 
•  Valve Maintenance......................page 4-8 
 
 
•  Valve Adjustment...................... page 4-12 
            •  Transducer Calibration .............. page 4-15 
Flow Path Maintenance 
The transducer flow path is 316 stainless steel (wetted magnetic 
parts of the solenoid valve are 430F stainless steel) with Viton
®
 
Neoprene
®
 or Kal-Rez
®
 (or equivalent) seals, depending on the 
gas used. The flow path should be periodically inspected and 
cleaned as required. 
 
When toxic or corrosive gases are used, purge unit thoroughly with 
inert dry gas before disconnecting from the gas line. A transducer 
used with toxic or corrosive gas must NOT be returned to the fac-
tory. Contact Technical Support for details. 
Cleaning the Laminar Flow Element 
The laminar flow element (LFE) is a precision flow divider which 
diverts a preset amount of flow through the sensor tube. The LFE 
is made of precision machined 316 stainless steel. The particular 
LFE used depends on the gas and flow range of the instrument and 
is identified by the number on the transducer nameplate. Should 
the LFE require cleaning or inspection due to deposition, use the 
appropriate cleaning procedure given on the following pages. 
 
  CAUTION! : REMOVING THE LFE WILL SHIFT THE 
CALIBRATION OF THE INSTRUMENT. DO NOT REMOVE THE 
LFE
 
UNLESS YOU PLAN ON RE-CALIBRATING THE TRANSDUCER.
 
 
 
Caution! 
It is important that this 
transducer be serviced 
only by qualified personnel. 
 
Caution! 
Always fully neutralize 
any toxic gas trapped in-
side the transducer 
before removing from 
the gas line.