Sierra 840 User Manual

Page of 74
Series 830/840/860 Instruction Manual 
Chapter 5 Troubleshooting
 
IM-83/84/86-H 
5-1
 
Chapter 5  Troubleshooting 
This chapter is provided to help locate the cause of a transducer 
failure. It is not intended to be an all inclusive repair manual. In 
the case of major repairs, the unit should be returned to the fac-
tory for service. Contact Sierra Customer Service for return 
shipping instructions (page 5-4). 
 
When you suspect that the transducer is not operating correctly, 
there are a few simple checks that can be made before dismantling 
the unit: 
 
1.  Make certain that there are no leaks in the line. 
 
2.  Check that all cables are connected and are in good condition. 
 
3.  Verify that the power supply is of the correct polarity and 
voltage and properly connected to the transducer. 
 
4.  For controllers, check for adequate pressure differential across 
the controller. Check the valve voltage. 
 
5.  Double check connector pin outs when replacing another 
manufacturer’s transducer. 
 
After verifying the factors above, follow the troubleshooting proce-
dures outlined on the next page. 
 
 
 
 
Warning! 
Always remove main power 
before disassembling 
the transducer. 
 
 
 
 
Caution! 
Only qualified personnel 
should perform trouble-
shooting procedures.