Siemens FX300 User Manual
3
INSTALLATION
20
SITRANS FX300
www.siemens.com/flow
06/2009 - SFIDK.PS.050.F2.02 A5E02100423
Recommended installations for measurement of liquids
Recommended installations for measurement of liquids
Recommended installations for measurement of liquids
Recommended installations for measurement of liquids
CAUTION!
It is absolutely necessary to comply with the required inlet and outlet runs.
It is absolutely necessary to comply with the required inlet and outlet runs.
1 If the device is installed in a downpipe, a standpipe must be installed immediately after it.
2 Installing the device in an inclined standpipe.
3 Installing the device in a vertical standpipe.
4 Installing the device in the lower pipe bend.
2 Installing the device in an inclined standpipe.
3 Installing the device in a vertical standpipe.
4 Installing the device in the lower pipe bend.
1 Above a horizontal pipe
2 On a vertical pipe
2 On a vertical pipe
INFORMATION!
Depending on the installation position, you may have to rotate the display and/or the connection
housing.
Depending on the installation position, you may have to rotate the display and/or the connection
housing.
MA_Siemens_Sitrans_FX300_R02_en_PRT.book Page 20 Monday, June 22, 2009 8:17 AM