Display Panel
English  202052A
Ready
The ready screen displays the serial number of the 
SpotOn sensor and indicates the SpotOn sensor 
cable and sensor are properly attached to the SpotOn 
control unit. The SpotOn sensor is now ready to be 
attached to the patient.
Equilibration 
A flashing yellow temperature display indicates the 
sensor has been connected to the patient and to the 
control unit and is in the process of equilibration. 
The graph below the temperature output indicates 
the progress of equilibration. This process will take 
approximately three minutes. 
If you see a temperature in the upper left hand corner 
of the screen, that is the last recorded temperature that 
has been captured by the sensor on the patient’s head. 
The sensor captures and records a temperature every 
five minutes.
After the equilibration is complete, the patient’s body 
temperature is displayed on the SpotOn control unit 
in bold white numbers and can be automatically 
transmitted to the patient monitor via the monitor 
cable. 
Running
The running screen displays the deep tissue 
temperature of the patient in white. The trend graph 
at the bottom of the screen displays the last two hours 
of the patient’s temperature in five minute increments. 
The bars on the graph will begin filling from the right 
with the 36°C always visible. The blue bars indicate 
that the temperature dropped below 36°C (white 
indicates the temperature is above 36°C).
Control unit error
The SpotOn control unit error screen displays when 
there is a system error. Discontinue use of unit. 
Contact a biomedical technician.
Sensor error
The sensor error screen displays when there is a 
temperature sensor or cable error. See “Section 4: 