2      Getting Started 
          
Up and Running overview 
 
  
 
Getting Started 
 
Up and Running overview 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
Placing the QUESTempº 48N on the job site 
The QUESTempº 48N should be placed at a height of 3.5 feet (1.1m) for standing 
individuals or 2 feet (.6m) for seated individuals.  Tripod mounting is recommended 
to get the unit away from anything that might block radiant heat or airflow.  A 1/4"x 
20 threaded bushing on the bottom of the instrument allows mounting to a 
standard photographic tripod.  Do not stand close to the unit during sampling. 
 
Before data logging, allow ten minutes for the sensors readings to stabilize.
1.  Place the QUESTempº 48N in the work area in a safe location approximately 
3.5 feet off the ground. 
 
2.  Turn the unit On.  If the battery voltage displayed during the power-on 
sequence is less than or equal to 6.4 volts, replace or recharge the 
batteries. 
 
 
3.  Be aware that the sensors require 10 minutes to stabilize to a new 
environment. 
 
 
 
4.  In the main menu, View will be selected (an indicator arrow denotes the 
selected menu).  Press the I/O Enter key and the measurement screen 
will appear. 
 
 
5.  Press the Run/Stop key to begin datalogging.  Use the Arrow keys to 
set the display to the desired items.  
 
 
 
Getting started