Fluke 199 User Manual

Page of 116
Fluke 192/196/199
Users Manual
88
Dual Input Oscilloscope
Isolated Inputs A and B (Vertical)
Bandwidth, DC Coupled
FLUKE 199 ....................................... 200 MHz (-3 dB)
FLUKE 196 ....................................... 100 MHz (-3 dB)
FLUKE 192 ......................................... 60 MHz (-3 dB)
Lower Frequency Limit, AC Coupled
with 10:1 probe ...................................... <2 Hz (-3 dB)
direct (1:1) ............................................. <5 Hz (-3 dB)
Rise Time
FLUKE 199 ....................................................... 1.7 ns
FLUKE 196 ....................................................... 3.5 ns
FLUKE 192 ....................................................... 5.8 ns
Analog Bandwidth Limiters .............. 20 MHz and 10 kHz
Input Coupling ................................................... AC, DC
Polarity ................................................ Normal, Inverted
Sensitivity Ranges
with 10:1 probe ............................50 mV to 1000 V/div
direct (1:1) ...................................... 5 mV to 100 V/div
Trace Positioning Range ..............................
±
4 divisions
Input Impedance on BNC
DC Coupled ..................... 1 M
Ω
 (
±
1 %)//15 pF (
±
2 pF)
Max. Input Voltage
with 10:1 probe ...................................... 600 V CAT III
1000 V CAT II
direct (1:1) ............................................. 300 V CAT III
(For detailed specifications, see “Safety”)
Vertical Accuracy ....................
±
(1.5 % + 0.04 range/div)
Digitizer Resolution ................... 8 bits, separate digitizer
for each input
Horizontal
Maximum Time Base Speed:
FLUKE 199 .....................................................5 ns/div
FLUKE 196 .....................................................5 ns/div
FLUKE 192 ................................................... 10 ns/div
Minimum Time Base Speed (Scope Record) .... 2 min/div
Real Time Sampling Rate (for both inputs simultaneously)
FLUKE199:
5 ns to 2 
µ
s /div ................................. up to 2.5 GS/s
µ
s to 120 s/div .......................................... 20 MS/s
FLUKE 196:
5 ns to 2 
µ
s /div .................................... up to 1 GS/s
µ
s to 120 s/div .......................................... 20 MS/s
FLUKE 192
10 ns to 2 
µ
s /div .............................. up to 500 MS/s
µ
s to 120 s/div .......................................... 20 MS/s