Raman Filter Chamber 
Description:  
The Princeton Instruments Model IARC-446-070 is a filter chamber which 
permits use of Raman notch or edge filters with Acton series monochromators 
and spectrographs. The Raman Filter Chamber is designed to operate over the 
wavelength range of 633nm to 1100nm.  Although the unit is used primarily to 
collect the output of fiber optics, it may also be used with other sources of 
Raman scatter (e.g., focused sample image or direct collection of scatter from a 
sample positioned close to the chamber).  
The standard IARC-446-070 includes a lens chamber with a filter holder designed 
to accept a 1.750 inch outside diameter filter or a 1.750 inch outside diameter 
filter assembly. A 1inch filter adapter is provided and installed in the Raman Filter 
Chamber.  This adapter is designed to accept a 1 inch or 25mm outside diameter 
filter/ filter assembly.  Other filter holders are available on a custom order basis. Filters can be rotated up to 10° 
by an external set screw in order to tilt-tune the rejection band of the filter.  
Principle of Operation: 
 The IARC-446-070 is designed to provide a collimated beam 
for operation with notch or edge filters. The first lens collects 
and collimates the beam through the filter. The second lens 
collects the collimated beam and focuses it to the entrance slit 
of the monochromator or spectrograph. The light is then 
dispersed to a detection system.  
The integrated tilt adjustment capabilities of the IARC-446-070 
enable fine tuning of notch or edge filters to achieve the best 
combination of laser light blocking and Raman scatter 
transmission.  
 
 
 
 
Figure 1. Raman Notch/Edge  
Filter Chamber 
 
Figure 2. Inside View of Chamber Showing the  
Filter Holder between the Focusing and  
Collimating Lenses