Type
Specification
Firmware
Embedded firmware. Can be upgrade with MON2020.
Methods
4 Timed Event tables and 4 Component Data tables assignable to each stream.
Peak Integration
•
Fixed time or auto slope and peak identification.
•
Update retention time upon calibration or during analysis.
2.2.5
Corrosion protection
Type
Specification
Enclosure Material
Copper-free and aluminum-coated with indus-
trial grade powder coat suitable for high humidi-
ty and salt-laden environments.
trial grade powder coat suitable for high humidi-
ty and salt-laden environments.
Process Wetted Materials
Stainless steel; if the function of an item ex-
cludes the use of stainless steel, such as the
glass rotameter tubes, materials that are resist-
ant to corrosion are used.
cludes the use of stainless steel, such as the
glass rotameter tubes, materials that are resist-
ant to corrosion are used.
Electronics
All electronic circuit boards are tropicalized with
a clear conformal coating.
a clear conformal coating.
2.2.6
Archived Data Storage Capabilities
Type
Maximum Number of Re-
cords
cords
Remarks
Analysis Results
31744
88 days with 4 minute cycle
time
time
Final Calibration Results
370
1 year of Final Calibration re-
sults
sults
Calibration Results
100
Final Validation Results
370
1 year of Final Validation results
Validation Results
100
Analysis Chromatograms
1703
Approximately 4.5 days assum-
ing 4 minute cycle time
ing 4 minute cycle time
Final Calibration Chromato-
grams
grams
370
1 year of Final Calibration Chro-
matograms*
matograms*
Final Validation Chromato-
grams
grams
370
1 year of Final Validation Chro-
matograms*
matograms*
Protected Chromatograms
100
User -selectable
Hourly Averages (Up to 250**
variables)
variables)
2400
100 days
Equipment description and specifications
24