7-6  Correcting the Internal Mass Value 
The GP series balance can correct the internal mass value within the range shown below. This 
function corrects the internal mass value to conform to an external weight. The corrected mass value 
is maintained in non-volatile memory even if the AC adapter is removed. 
The internal mass value is corrected as follows: 
Model 
Correction reference value 
Correction range 
GP-12K 10 
kg 
GP-20K 
GP-30K/GP-30KS 
GP-32K/ GP-32KS 
GP-61K/ GP-61KS 
20 kg 
GP-40K 40 
kg 
 
 
-5.0 g to +5.0 g 
GP-60K/ GP-60KS 
60 kg 
-50 g to +50 g 
GP-100K/GP-100KS 
GP-102K 
100 kg 
-50 g to +50 g 
 
Example: Using the GP-20K