A&D GP-102K User Manual

Page of 98
 
 65
11-5  Memory for Comparator Settings 
 The data memory function can store 20 sets of upper and lower limit values for the comparator mode. 
The reference value or tolerance value for the comparator mode can not be stored in memory. 
  The upper and lower limit values in memory can be recalled easily using the MODE key and 
used for weighing. 
  The upper and lower limit values in memory can be recalled and changed. 
Note 
The recalled upper and lower limit values can be changed, using the digital input mode or the 
weighing input mode. The digital input mode enters the upper and lower limit values using 
the keys. The weighing input mode uses a sample to store the upper and lower limit values.   
 
Storing the upper and lower limit values 
To store new upper and lower limit values: Recall the stored upper and lower limit values to be 
changed (“C01” to “C20”). Then, change the recalled upper and lower limit values using the digital 
input mode or the weighing input mode, and store the new value. 
1  Press the MODE key to select a unit to be used for storage. 
Note 
While the data memory function is in use, unit selection using the MODE key is not available. 
2  Set the “Data memory (data)” parameter to “4”. 
 If 
“Clr” appears blinking in the upper left of the display, delete the data in memory as 
described in “11-1 Notes on Using Data Memory”.
 
3  Press and hold the PRINT key until the balance enters the upper and lower limit values 
confirmation mode. The upper limit value last selected is displayed as shown below.   
 
 
 
                                                                                                                All the digits blinking 
4  Select the comparator number to be used, using the following keys. 
RE-ZERO key 
To increase the comparator number by one.   
MODE key 
To decrease the comparator number by one. 
 
 Each time the RE-ZERO key or MODE key is pressed, the displayed 
value changes as follows: .... 
↔ C03 HI ↔ C03 LO ↔ C04 HI ↔ C04 LO 
↔ .... 
5  To change the selected upper and lower limit values: 
To use the digital input mode, press the SAMPLE key to enter the digital input mode. Go to 
“Digital input mode” on the next page. 
To use the weighing input mode, press the SAMPLE key, then press and hold the MODE key 
to enter the weighing input mode. Go to “Weighing input mode” on the next page. 
Note 
Using the “CN:mm” command, the upper and lower limit values can be recalled. (mm 
indicates a two-digit numerical value 01-20, which corresponds to C01-C20.) The upper and 
lower limit values recalled can be output using the “?HI” command or “?LO” command and 
can be changed using the “HI:” command or “LO:” command.