  The following commands can not be used during data storage. 
 
Query command for weighing data.   
 
Query command for stable weighing data. 
 
SI 
Query command for weighing data.   
 
SIR 
Query command for continuous weighing data. 
 
Setting the function table   
Parameter settings for each output mode are as follows: 
                            Item 
 
Mode 
Data output 
mode 
Auto print 
polarity, 
difference 
Data memory 
function 
Interval 
time 
Key mode 
prt 0 
Not used 
data 2 
Auto print mode A 
prt 1 
data 2 
Auto print mode B 
prt 2 
ap-a 0-2 
ap-b 0-2 
data 2 
 
Not used 
Interval memory mode 
prt 3 
Not used 
data 2 
int 0-8 
 
Set each item, depending on the situation, as follows:   
No 
d-no 0 
No 
5-td 0 
Data number   
Yes 
d-no 1 
Time only 
5-td 1 
No  
5-id 0 
Date only 
5-td 2 
ID number 
Yes 
5-id 1 
 
Time and date 
Both 
5-td 3 
    With 5-td 1, 5-td 2 or 5-td 3 selected, the amount of data to be stored is 100 sets. 
 
Enabling the data memory function 
1  Press and hold the SAMPLE key until    ba5fnc   is displayed, then release the key. 
2  Press the SAMPLE key several times to display   dout . 
3  Press the PRINT key. 
4  Press the SAMPLE key three times to display   data 0 . 
5  Press the RE-ZERO key to display   data 2 . 
6  Press the PRINT key to store the setting. 
7  Press the CAL key to return to the weighing mode.