11-4  Memory for Unit Mass in the Counting Mode 
  The data memory function can store 50 sets of unit mass for the counting mode. Even if the 
AC adapter is removed, the data is maintained in non-volatile memory. Among the 50 sets, 
“p01” is the memory function for the standard counting mode. 
  The unit mass in memory can be recalled and used for weighing. 
  The unit mass in memory can be recalled and changed. 
Note 
The recalled unit mass can be changed, using the weighing input mode or the digital input 
mode. The weighing input mode uses the specified number of samples to store the unit mass. 
The digital input mode enters the unit mass using the keys. 
 
Storing the unit mass 
To store a new unit mass: Recall the stored unit mass to be changed (“Pp01” to “p50”). Then, change the 
recalled unit mass using the weighing input mode or the digital input mode, and store the new unit mass. 
1  Set the “Data memory ( data )” parameter to “1”. 
2  Press the MODE key to select        pcs    (counting mode).   
 If 
“Clr” appears blinking in the upper left of the display, delete the data in memory. 
Note 
If the counting mode can not be selected, refer to “4. WEIGHING UNITS”. 
3  Press and hold the PRINT key until the balance enters the sample unit mass confirmation 
mode. The unit mass last selected is displayed as shown below.   
 
 
                 
                                                                                          All the digits blinking 
4  Select the unit mass number to be used, using the following keys. 
RE-ZERO key 
To increase the unit mass number by one.   
MODE key 
To decrease the unit mass by one. 
5  To change the selected unit mass: 
  To use the weighing input mode, press the SAMPLE key to enter the weighing input mode.   
Go to “Weighing input mode” on the next page. 
  To use the digital input mode, press the SAMPLE key, then press and hold the MODE key to 
enter the digital input mode. Go to “Digital input mode” on the next page. 
Notes 
ACAI can not be performed directly on the recalled unit mass. 
Using the “UN:mm” command, the unit mass can be recalled. (mm indicates a two-digit 
numerical value 01-50, which corresponds to P01-P50.) The unit mass recalled can be output 
using the “?UW” command and can be changed using the “UW:” command.