8-2  Initializing the Balance 
This function returns the following parameters to factory settings. 
 Calibration data 
 Function table 
  The sample unit mass value (counting mode), 100% reference mass value (percent mode) 
  The data that is stored in the balance using the data memory function 
  External calibration weight and target weight value 
  Function switch settings 
  Liquid density and temperature in the density mode 
 
Note 
Be sure to calibrate the balance after initialization. 
 
Operation 
1  Press the ON:OFF key to turn off the display. 
2  While pressing and holding the PRINT key and the  
SAMPLE key, press the ON:OFF key. The balance 
displays  p5  . 
3  Press the SAMPLE key to display    Clr . 
4  Press the PRINT key.   
  To cancel this operation, press the CAL key. 
5  Press the RE-ZERO key. 
6  Press the PRINT key to initialize the balance. 
  The balance will automatically return to the weighing 
mode.