A&D GP-100K User Manual

Page of 98
 
 43
Interval memory mode 
The weighing data is periodically stored in memory.   
Required setting 
dout prt 
Interval memory mode 
 
dout data 
Data memory function is used. 
 
 
 
Stores weighing data. 
 
dout int Interval 
time 
Optional setting 
dout  
5-td1, 2, or 3  Adds the time and date. 
Example 
“For periodical weighing without a computer command and to output 
all of the data, to a computer, at one time” 
 
9-6  Description of the Item “Data format” 
A&D standard format   
5if type 0  
This format is used when the peripheral equipment can receive the A&D format. If an AD-8121B is 
used, set the printer to MODE 1 or 2. 
  This format consists of fifteen characters excluding the terminator. 
  A header of two characters indicates the balance condition. 
  The polarity sign is placed before the data with the leading zeros. If the data is zero, the plus 
sign is used.   
  The unit, consisting of three characters, follows the data. 
 
 
 
 
 
 
DP (Dump print) format 
5if type 1 
This format is used when the peripheral equipment can not receive the A&D format. If an AD-8121B 
is used, set the printer to MODE 3. 
  This format consists of sixteen characters excluding the terminator. 
  A header of two characters indicates the balance condition. No overload header is used. 
  The polarity sign is placed before the data, with spaces in place of leading zeros, if the data is 
not zero or overloaded.   