14.  I/O UNIT SPECIFICATIONS 
(Standard)
 
14-1 RS-232C//External Contact Input  
D-Sub 25 pin numbers 
 
 
 
 
 
 
D-Sub 25 pin assignments 
 
 
Pin No.  Signal name 
Interface type 
Direction
Description 
1 FG 
 
Frame 
ground 
2 RXD 
RS-232C 
Input 
Receive 
data 
3 TXD 
RS-232C 
Output 
Transmit 
data 
RTS 
RS-232C 
Input 
Ready to send 
CTS 
RS-232C 
Output 
Clear to send 
DSR 
RS-232C 
Output 
Data set ready 
SG 
RS-232C/external contact input
Signal ground 
18 
PRINT 
External contact input 
Input 
Same as the PRINT key 
19 
RE-ZERO 
External contact input 
Input 
Same as the RE-ZERO key
Others - 
No 
connection 
 
 
RS-232C 
The balance is a DCE device. Connect the balance to a personal computer (DTE), using a straight 
through cable. 
Transmission system 
:  EIA RS-232C 
Transmission form 
:    Asynchronous, bi-directional, half duplex 
Transmission rate 
:  10 times/second or 5 times/second (same as data refresh rate) 
Data format 
:    Baud rate 
:  600, 1200, 2400, 4800, 9600, 19200bps 
 
    Data bits   
:  7 or 8 bits 
 
  Parity 
:  Even, Odd    (Data bits 7 bits) 
 
   
  None   
(Data bits 8 bits) 
 
    Stop bit 
:  1 bit     
(When sending, 2 bits; receiving, 1 bit.         
A personal computer will function with either 
setting.) 
  
Code 
ASCII