Teledyne T201 User Manual

Page of 44
Teledyne API Model T201 NH
Analyzer Operator Manual
 
T201 Ammonia Analyzer
 
07271B DCN6646 
13 
1.4  T201
 
A
NALYZER 
S
PECIFICATIONS
 
Table 1-1. T201 Operating Specifications  
Ranges 
Min: 0-50 ppb Full scale 
Max: 0-2000 ppb Full scale (selectable, independent NH
3
, NO, NO
2
NOx ranges supported) 
Measurement Units 
PPB, PPM, microgram/m
3
, milligram/m
(user selectable) 
Noise at Zero
1
 
< 0.5 ppb RMS 
Noise at Span
1
 
< 1.0% of reading above 50 ppb 
Lower Detectable Limit
1 ppb RMS 
Zero Drift
3, 6
 
2 ppb / 24 hours  
Span Drift
6
 
< 1.0% FS  Range / 24 hours   
Lag Time 
40 seconds 
Rise Time
90% 300 seconds 
Fall Time
4
 
90% 300 seconds 
Sample Flow Rate 
1000 cm
3
/min + 10% (500 cm
3
/min bypass to vacuum manifold, 500 
cm
3
/min to reaction cell) 
Linearity 
NO calibration  1% of full scale 
NH
3
 calibration  2% of full scale 
Temp Range 
15-40 
o
Dimensions H x W x D 
7” x 17” x 23.6” (18cm x 43cm x 61cm) 
Weight, Analyzer 
43 lbs (20 kg) 
Weight, Converter 
24 lbs (11 kg) 
Weight, Pump 
16 lbs (7 kg) 
Power, Analyzer 
100V ~50/60 Hz, 120V ~60 Hz, 220V ~50Hz, 240V ~50 Hz, 125 watts 
Power, Analyzer
5
 
230V ~50 Hz, 125 watts 
Power, Pump 
110V ~60 Hz, 220V ~50 Hz, 240V ~50 Hz, 295 watts 
Power, Pump CE Mark
5
 
230 V ~50 Hz, 2.5 A peak 
Environmental 
Installation Category Pollution Degree 2, Over-voltage Category II 
Analog Output Resolution  
1 part in 4096 of selected full-scale voltage (12 bit) 
standard 
(1) Ethernet: 10/100Base-T; (2) RS-232 (300 – 115,200 baud)  
(2) USB device ports; (8) opto-isolated digital outputs; (6) opto-isolated 
digital inputs; (4) analog outputs 
Serial I/O 
optional 
(1) USB com port; (1) RS485 (300 – 115,200 baud); (8) analog inputs 
(0-10V, 12-bit); (4) digital alarm outputs; Multidrop RS232; (3) 4-20mA 
current outputs 
1
  As defined by USEPA 
2
  Defined as twice the zero noise level 
3
  At constant temperature and voltage 
4
  When pneumatics are conditioned with NH
overnight and the gases (zero air to NH
3
) are manually 
switched at the sample inlet at the rear of the M501. 
5
  Electrical rating for CE Mark Compliance 
6
  Applies when sampling NH
3