Tektronix AM700 User Manual
ROUTe Subsystem
3–102
AM700 Audio Measurement Set Programmer Manual
ROUTe Subsystem
Signal routing is the block where the user has access to actual signals.
ROUTe[1–3]:CLOSe <numeric_value>
Usage:
Sets the route setting. If all the specified channels cannot be closed, an
execution error is reported.
execution error is reported.
Suffixes:
The ROUTe suffixes are as follows:
1
Analog A input
2
Analog B input
3
Digital main input
Parameters:
Defined values for the CLOSe command <numeric_value>
1
analog A input connector
2
analog B input connector
3
analog generator channel A
4
analog generator channel B
6
digital front panel XLR
7
digital rear panel BNC
8
digital rear panel optical
9
AES internal generator
Only specific numeric_values are permitted for each route suffix. You can’t close a
route from the Analog A input to the Digital front panel XLR for instance. The
acceptable routes and closures are as follows:
route from the Analog A input to the Digital front panel XLR for instance. The
acceptable routes and closures are as follows:
rout1:clos 1|3
rout2:clos 2|4
rout3:clos 6|7|8|9
Example:
Choose the generator for analog B input as follows:
ROUTe2:CLOSe 4
Query:
ROUTe[1–3]:CLOSe? <channel_number>