12-8
 (E)
SIR-3400H / SIR-3032i / SIR-3032W
12-8. Setting the Transmission Format
Set the S/D select switch on the front of measuring unit to the desired format.
1
When using the IEPE input, set to S.  When set to D, the unit does not work.
Setting
Transmission format
Supported input format
S
Single ended input
Direct input, IEPE input
D
Differential input
Direct input
When using in the differential input, connect the signal source to the measuring unit as shown below.
Unless the signal source is connected with the frame ground of measuring unit, signal may not be
properly recorded.
Signal source
Measuring unit
Signal hot
– BNC terminal (Core of coaxial cable)
Signal cold
– BNC GND (Shield wire of coaxial cable)
Ground terminal – Ground terminal (Frame ground connection)
12-9. Recording Memo 
(Modes: STOP, E-E, REC-PAUSE, and REC)
Up to 96 characters of information can be set during recording.
1
Press the MEMO key.
The lamp at the top left of the key turns on, and the Record Memo indicator appears in the setting value
display area to indicate that the memo can be edited.
12. Operation Settings (Using Setting key)