Important EMC notices for use in the medical environments
• The UP-DR80MD needs special precautions regarding 
EMC and needs to be installed and put into service 
according to the EMC information provided in this 
instructions for use.
• The portable and mobile RF communications equipment 
such as cellular phones can affect the UP-DR80MD.
Warning
The use of accessories and cables other than those 
specified, with the exception of replacement parts sold by 
Sony Corporation, may result in increased emissions or 
decreased immunity of the UP-DR80MD.
Warning
If the UP-DR80MD should be used adjacent to or stacked 
with other equipment, it should be observed to verify 
normal operation in the configuration in which it will be 
used.
Guidance and manufacturer’s declaration-electromagnetic emissions
The UP-DR80MD is intended for use in the electromagnetic environment specified below.
The customer or the user of the UP-DR80MD should assure that it is used in such an environment.
Emission test
Compliance
Electromagnetic environment-guidance
RF emissions
CISPR 11
Group 1
The UP-DR80MD uses RF energy only for its internal 
function. Therefore, its RF emissions are very low and are 
not likely to cause any interference in nearby electronic 
equipment.
RF emissions
CISPR 11
Class B
The UP-DR80MD is suitable for use in all establishments, 
including domestic establishments and those directly 
connected to the public low-voltage power supply network 
that supplies buildings used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Complies