 125 − 
 
6 F 8 A 0 9 3 4  
10.8 Converter Failure Alarm 
If any one of the following errors occurs in a self-diagnosis sequence, the converter issues an alarm 
using a contact output. 
 
●Self-diagnosis errors 
Self-diagnosis errors (LCD display) 
Error items 
ROM ERROR 
ROM error 
RAM ERROR 
RAM error 
PARAMETER 
    FAILURE 
System parameter error 
EXC CUR 
       OPEN 
Excitation circuit not connected or open 
EXC CUR 
      ERROR 
Excitation current error, excitation circuit error 
ADC ERROR 
ADC error 
INVALID 
      TOTAL 
Invalid totalizer counts 
Note: A self-diagnosis error message appears on the Display 2 measured value screen. If this 
happens, the measurement item specified on the Display 2 screen cannot be displayed 
unless the error is removed. However, if OFF is set to 8.2.24 
, “Self Diagnosis Function Setting,” an error does not appears even if an error occurs. 
 
If you want to use a converter alarm output, set Digital Output 1 (DO1) or Digital Output 2 
(DO2) for converter alarm output (CONV ALM) following 8.2.18, “Digital Output.” 
In addition, set the alarm output condition to normally open (NormOPEN) or normally 
closed (NormCLOSE) status. 
 
●Contact output condition 
 
NormOpen;     
In case an error occurs, contact is closed. 
NormClose; 
In case an error occurs, contact is open.   
 
Note:    Alarm output contact is open when converter power is off.