Emerson NGA2000 User Manual

Page of 178
4.2.1  Calibration - 
Zeroing
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
NGA 2000
4 - 23
MEASURE
STATUS
NEXT
LOCK
INFO
CLD
F1
F2
F3
F4
F5
0.017 ppm
Main Menu
Basic controls...
Expert controls and set up...
Technical level configuration...
(Operational configuration)
(Diagnostic and manufacturing/service)
DISPLAY
STATUS
MENU
NEXT
INFO
CLD
Sample press.:
Detector temperature:
Block temperature:
Converter temp.:
51.5 C
0.5 C
340 hPa
360 C
F1
F2
F3
F4
F5
0.017
ppm NOx
Range: 4
0
250
⇒ 
⇒ 
Change to the
single component
display
Press  
F1
⇒ 
⇒ 
Single component
display will appear
after the zero gas
calibration
⇒ 
⇒ 
Notes for zero gas calibration:
• 
You can calibrate all ranges together or each range separately
with zero gas.
• 
You have to set up the parameter you want in the menu
"Calibration Parameters" (see chapter 5.1.2 S. 5-9).