Agilent Technologies 664xA User Manual

Page of 70
 52     Status Reporting
 Table 4-2.  Bit Configurations of Status Registers
 Bit
 Signal
 Meaning
 
 Bit
 Signal
 Meaning
 
 
 Operation Status Group
 
 
 
 
 Standard Event Status Group
 0
 CAL
 The interface is computing new
calibration constants.
 
 0
 OPC
 Operation complete.
 5
 WTG
 The interface is waiting for a trigger.
 
 2
 QYE
 Query error.
 8
 CV
 The power module is in constant
voltage mode.
 
 3
 DDE
 Device-dependent error.
 10
 CC
 The power module is in constant
 
 4
 EXE
 Execution error.
 
 
 current mode.
 
 5
 CME
 Command error.
 
 
 
 
 7
 PON
 Power on.
 
 
 Questionable Status Group
 
 
 
 
 Status Byte and Service Request
Enable Registers
 
 0
 OV
 The power module overvoltage
protection circuit has tripped.
 
 3
 QUES
 Questionable status summary bit.
 1
 OC
 The power module overcurrent
protection circuit has tripped.
 
 4
 MAV
 Message Available summary bit.
 4
 OT
 The power module has an
overtemperature condition.
 
 5
 ESB
 Event Status summary bit.
 9
 RI
 The power module remote inhibit state
is active.
 
 6
 MSS
 RQS
 Master Status summary bit.
 Request Service bit.
 10
 UNR
 The power module output is
unregulated.
 
 7
 OPER
 Operation status summary bit.
 
 
 Figure 4-1. Power Supply Status Model