Tektronix AFG3000 User Manual

Page of 230
Syntax and Commands
AFG3000 Series Arbitrary/Function Generators Reference Manual
3-31
*OPC
This command generates the operation complete message by setting bit 0 in the 
Standard Event Status Register (SESR) when all pending commands that generate 
an OPC message are complete.
The query command places the ASCII character “1” into the output queue when all 
such OPC commands are complete.
Group
Synchronization
Syntax
*OPC
*OPC?
Arguments
None
Returns
<execution complete>::=1
where “1” indicates that all pending operations are complete.
Examples
*OPC?
might return 1 to indicate that all pending OPC operations are finished.
*OPT?
This query-only command returns a list of the options installed in your arbi-
trary/function generator.
Group
System
Syntax
*OPT?
Arguments
None
Returns
<OPT>[,<OPT>[,<OPT>[,<OPT>]]]
Examples
*OPT?
might return 0, which indicates no option is installed in the instrument.