Syntax and Commands
3-86
AFG3000 Series Arbitrary/Function Generators Reference Manual
[SOURce[1|2]]:VOLTage:LIMit:HIGH
This command sets or queries the higher limit of the output amplitude high level 
for the specified channel. If your instrument is a dual-channel model and the 
[SOURce[1|2]]:VOLTage:CONCurrent[:STATe] command is set to ON, then the 
higher level limit of the other channel is the same value.
Group
Source
Related Commands
[SOURce[1|2]]:VOLTage:CONCurrent[:STATe]
Syntax
[SOURce[1|2]]:VOLTage:LIMit:HIGH {<voltage>|MINimum|MAXimum}
[SOURce[1|2]]:VOLTage:LIMit:HIGH? {MINimum|MAXimum}
Arguments
<voltage>::=<NRf>[<units>]
where
<NRf> is the higher limit of output amplitude.
<units>::=[mV | V]
Returns
<voltage>
Examples
SOURce1:VOLTage:LIMit:HIGH 1V