6.8 
ANIMAL
 
(D
YNAMIC
)
 
WEIGHING
 
 
The scale can  be  set to  animal  (dynamic) weighing for weighing  any items that are 
unstable or may move.   See Section 7.4. 
 
The scale will use a special filter to minimise the effects of any movement while the 
moving animal or unstable samples are on the scale.  
  
•  Press  [Func/C].  The  first  option  is  “FunC 1”,  press  the  [Func/C]  key  again  to 
display “Func 2”.  
 
 
•  Press  the  [Z/T]  key.  “F2  PCt”  will  be  displayed.  Press  the  [Pcs/
]  key  to 
advance to the second function, “P2 AnL”,   Animal weighing. 
 
 
•  Press [Z/T] to enter animal weighing. The scale will change to animal weighing 
and allow weighing of an unstable animal or sample on the platform.  
 
•  To  use  the  Animal  Weighing  function  it  is  necessary  to  set  the  amount  of 
filtering  required  for  the  item  to  be  weighed.      More  active  animals  will 
require a higher level of filtering to give a stable result.    The display will show 
“Flt x”  where  x  is  a  value  from  1  to  5.    The  higher  the  value  the  greater  the 
amount of filter will be.  To increment the value shown press the [Pcs/
] key 
then press the [Z/T] key to accept it. 
 
 
•  The display will flash “Ani   “ 2 times then show the current weight, 0.00.  The 
scale is now ready to weigh.