Adam Equipment 3106610631 User Manual

Page of 44
35 | 
P a g e  
 
 
 
 
 
 
© Adam Equipment Company 2010
 
 
 
12.3.3  
F3 – CAPACITY/ WEIGHING UNIT/INCREMENT 
 
To enter this parameter, press the [Z/T] key when “F3 CAP” is shown. 
The display will show the current capacity and default weighing unit.   
Enter  the  numeric  values  using  the  [Unit/
]  and  [Pcs/
]  keys.        The  scale  will 
check the number of division’s n = maximum/increment is less than 30,000 divisions.  
(3000 divisions for EC approved versions).  
 
To  select  the  weighing  unit  press  the  [Unit/
]until  the  default  weighing  unit  is 
flashing, either kilogram (kg) or pounds (lb).    
 
Press [Z/T] to accept the displayed value. 
 
On non-approved versions the display then lets you select the increment, “InC 2” 
For example 100kg x 0.01kg the increment is 10grams, but the last digit increments 
by 1. 
 
The display will show the current increment value as used with the current decimal 
point position. 
 
Press the [Pcs/
] key to change the value. Select from 1, 2, 5, 10, 20 or 50 
 
Not all increments may be available for the capacity you selected. 
 
For EC Approved versions the scale will determine the increment that maintains the 
number of divisions to be 3000 or less. 
 
Press [Z/T] to accept the displayed value. 
Press [Print/M+/Esc] to return to weighing.