Operator’s Checkout Procedure
1760 Series Component Waveform/Vector Monitor User Manual
2-13
Procedure
1.
Initialize the Front-Panel Controls
Enter the
PRESET
menu and select
FACTORY
. Press the bezel
button corresponding to
RECALL
. The front-panel controls
are now set to the factory preset (described on page 3-18),
and the
PRESET
menu is automatically exited. Notice that
the instrument is in the
WAVEFORM
display mode with the
CH-A INPUT
displayed in two-line sweep.
2.
Adjust the Display
Use the
VERT POS
and
HORIZ POS
controls to center the
display.
Enter the
CRT
menu.
DISPLAY
will be outlined. Adjust the
bezel controls for optimum focus, graticule scale illumina-
tion, and signal intensity. Select
READOUT
and adjust the
right bezel control for desired readout intensity. Select
TRACE
and adjust the center bezel control for a level trace.
Exit the
CRT
menu. The changes made while in the menu
remain in effect.
3.
Input Channel Selection
Press the
CH-A/CH-B
button to toggle from the
CH-A
input to
the
CH-B
input (no signal applied).
Press the
CH-A1/CH-B1
button. This cancels the
CH-B
input
and displays the
CH-A1
input. Press this button again to
toggle to the
CH-B1
input.
Select the Y, B--Y, R--Y output format on the signal genera-
tor. Press the 1760--Series
A123/B123
button. This cancels
the
CH-B1
input. The
A123
indicator lights and a component
parade of
CH-A1, CH-A2
, and
CH-A3
is displayed. See
Figure 2-5. Press the
A123/B123
button again; this toggles to
B123
(a display of the
CH-B1, CH-B2
, and
CH-B3
inputs).
Select any other input, which cancels the
B123
selection.
Now press the
A123/B123
button again, and note that the
B123
selection is restored.