Adam Equipment qbw User Manual

Page of 16
© Adam Equipment Company 2004 
 
 
  If the no-load calibration value is acceptable, the display will then show 
""
 where 
is the value of calibration mass required and “
""
” 
is the units (kg or lb) for the calibration mass, e.g. 
 for a 6 kg 
mass or “
” for a 10 pound mass. 
 
 
 
  Place the correct calibration mass on the scale. 
 
  Press the  [Cal] key again. The display shows “
” until calibration is 
complete and then returns to normal weighing.  
 
  Verify the calibration by placing the calibration mass on the scale. The 
display  should  show  the  correct  calibration  mass  in  the  currently 
selected units. 
 
NOTE: 
  If the no-load or full-load reading is too low 
  will be displayed  
  If the no-load or full-load reading is too high 
 will be displayed 
  If the no-load or full-load reading is unstable 
 
 will be displayed.        
             
If any of these conditions occurs, or if the [Cal] key is not pressed within 10 
seconds of the “
” message being displayed, the calibration procedure will 
be  aborted  and  the  scale  will  return  to  normal  weighing  using  the  previous 
tare and calibration values. 
 
6.0  USER SET UP PARAMETERS 
        
 
6.1  FUNCTION OF THE KEYS 
 
To enter the user-parameter setup menu, the [Zero/Tare] key should be held 
pressed, and  while the display is showing 
 the [Units] key should be 
pressed. The first setup option will then be shown. Release both the keys. If 
[Zero/Tare]  is  released  without  the  [Units]  key  being  pressed,  the  display 
will  change  to 
,  a  normal  tare  operation  will  be  performed,  and  the 
setup menu will not be entered. 
 
When in the setup menu, the functions of the keys are as follows: 
 
  Press [Zero/Tare] to select alternatives within any parameter.