7.0  CALIBRATION
 
 
  Calibration is recommended for best accuracy. It is suggested that the 
scale be calibrated when first installed and any time the scale is moved 
to a different environment.  
   To ensure best accuracy calibrate before use.  
  Ensure weights are ready before calibrating, the weight should be of a 
known accuracy,  for example OIML class M1 or Class 4 of ASTM E617.  
  Do not calibrate the scale if a proper calibration weight is available.  
  Calibration can only be accessed if the Calibration function has been 
enabled as described in section 6.0. 
 
 
7.1 
AUTOMATIC
 
EXTERNAL
 
CALIBRATION
 
 
 
  In the weighing mode, press the [Func/
] key to access calibration 
mode. The display will show “CAL”. 
  Press the  [Tare/↵
] key, an access code (Pin code)  is required to 
continue,  The display will show “Pin” . 
  Press the [Func/
] key then [Units/
] key then [Tare/↵
] key, and 
then press [Zero/
ESC
] key to confirm the pin code.  
  The display will show the weighing unit last used to calibrate the scale, 
“S-Unit  kg”(or ”S-Unit  Lb”) ,press the [Hold/
] key to change to lb’s or 
kg’s for the calibration weighing unit value. 
  Press the [Tare/↵
] key to start the zero calibration. The display will 
show “CAL 0” and when the stable symbol is displayed press the [Tare/↵
key to confirm.  
  The display will now show the calibration mass expected.  ”CAL  XX Kg 
or ”CAL  XX Lb “
.Press the [Hold/
] key to change the calibration 
values if necessary. Each model has 3 different kg and lb calibration 
selection values.