Adam Equipment azextra-p User Manual

Page of 24
 
© Adam Equipment Company 2011 
17 
 
4.0  ERROR CODES 
 
During the initial power-on testing or during operation, the scale may show an 
error  message.    If  an  error  message  is  shown,  repeat  the  step  that  caused 
the message. If the error is still shown contact your dealer for further support.   
 
 
ERROR   DESCRIPTION 
POSSIBLE CAUSES 
Err  1 
Time input error. 
Tried to set an illegal time, i.e. 
26 hours  
 
Err  2 
Date input error 
Tried to set an illegal date, i.e. 
36
th
 day  
 
Err  4 
Initial Zero is greater than 
allowed (typically 4% of 
the maximum capacity) 
when power is turned on 
or when [Zero] is 
pressed, 
Weight is on the pan when 
turning the scale on. 
Excessive weight on the pan 
when zeroing scale. 
Improper calibration of the 
scale. 
Damaged load cell. 
Damaged Electronics. 
 
Err 6 
A/D count is not correct 
when turning the scale 
on. 
Platform is not installed. 
Load cell is damaged. 
Electronics is damaged.