11.0 ERROR CODES 
 
During  the  initial  power-on  testing  or  during  the  operation,  the  scale  may 
show an error message.  The meaning of the error messages are described 
below. 
 
If  an  error  message  is  shown,  repeat  the  procedure  that  caused  the 
message, such as turning the balance on, calibration or any other functions.  
If  the  error  message  is  still  shown  then  contact  your  dealer  for  further 
support.   
 
ERROR CODE 
DESCRIPTION 
POSSIBLE CAUSES 
Err  4 
Initial Zero is greater than the 
permissible value (typically 4% 
of maximum capacity) when 
power is turned on or when the 
[Zero] key is pressed, 
Weight on the pan when turning 
the scale on. 
 
Excessive weight on the pan 
when zeroing the scale. 
 
Improper calibration of the scale. 
 
Damaged load cell. 
 
Damaged Electronics. 
 
Err 5 
Keyboard error.   
Improper operation of the scale. 
 
Err 6 
A/D count is not correct when 
turning the scale on. 
Platform is not installed. 
 
Load cell may be damaged. 
 
Electronics may be damaged. 
 
FAIL HiGH 
or
  
FAIL LoW 
Calibration error 
Improper calibration. 
 
If  the  problem  persists  contact 
your  dealer  or  Adam  Equipment 
for assistance.