14.0 ERROR CODES 
 
During the initial power-on testing or during operation, the scale may show an 
error message.  The meaning of the error messages is described below. 
 
If  an  error  message  is  shown,  repeat  the  step  that  caused  the  message.  If 
the error message is still shown then contact your dealer for support. 
  
 
ERROR 
CODE 
DESCRIPTION 
POSSIBLE CAUSES 
 
Err  4 
Initial Zero is greater than 
allowed (4% of maximum 
capacity) when power is 
turned on or when [Zero] is 
pressed. 
Weight on the pan when turning the scale 
on. 
Excessive weight on the pan when zeroing 
the scale. 
Platform is not installed. 
Improper calibration of the scale. 
Damaged load cell. 
Damaged Electronics. 
Err 6 
A/D count is not correct 
when turning the scale on. 
Load cell is damaged. 
Electronics is damaged. 
Err  8 
High limit input error 
Low limit is set first, then the high limit is set 
lower than the low limit and high limit not 
equal to zero. 
Err 9 
Low limit input error 
High limit is set first, then the low limit is set 
higher than the high limit and low limit not 
equal to zero. 
FAIL H 
or
  
FAIL L 
Calibration error 
Improper calibration (should be within +10% 
of the factory calibration). The old calibration 
data will be retained until the calibration 
process is complete.