Anritsu MG3641A User Manual

Page of 212
SECTION 7   PERFORMANCE TEST
7-2
Power meter
7.2
Performance Test Device List
Table 7-1 lists the measuring instruments for calibration according to the test parameters.
Table 7-1.  Performance Test Device List
Test parameter
Measuring instrument
    Requested
Recommended
    performance*
device
Output frequency
Frequency counter
100 kHz to 2.08 GHz
MF1603A
Output level frequency
100 kHz to 2.08 GHz
ML4803A
characteristics
–30 to +20 dBm
MA4601A
Spectrum Analyzer
100 kHz to 2.08 GHz
MS2602A
Output level accuracy
Pre-amplifier
100 kHz to 2.08 GHz
40 dB
FM frequency deviation
Modulation analyzer
100 kHz to 2.08 GHz
MS616B
AM modulation factor
AM, FM
Modulation distortion
Distortion meter
20 Hz to 100 kHz
* Extracts a part of performance that can cover the test parameter measurement range.