Anritsu MG3641A User Manual

Page of 212
SECTION 7   PERFORMANCE TEST
7-5
7.3.2
Output level frequency characteristics
(1) Test specifications
• Frequency characteristics
with reference to 0 dBm
±0.5 dB
(2) Measuring instrument for test
• Power meter  (100 kHz to 2080 MHz)
(3) Setup
Fig. 7-2.  Output Level Frequency Characteristic Test
(4) Test procedure
STEP
                                                          PROCEDURE
1.
Adjust the power meter to the zero point and correct the sensor sensitivity.
2.
Set the MG3641A/MG3642A output level to 0 dBm.
3.
Set the MG3641A/MG3642A to the frequency to be measured.
4.
Set the calibration factor of the sensor of the power meter and read the output level.
5.
Repeat steps 3 and 4.
(5) Note on test
The MG3641A/MG3642A output level is defined with the RF output connector.  To measure the output level,
directly connect the power sensor to the MG3641A/MG3642A RF output connector.
Sensor
Power meter
SENSOR
INPUT
MG3641A/MG3642A