Anritsu MG3641A User Manual

Page of 212
SECTION 8   CALIBRATION
8-2
8.3
Calibration
For the MG3641A/MG3642A, calibrate the frequency of the built-in Reference oscillator  once a half year.
The stability of the MG3641A/MG3642A 10 MHz reference crystal oscillator is 
±5 × 10
-9
 per day.  Therefore, as the
frequency standard, use a standard signal generator that receives the typical radio wave and subcarrier of the color
telecasting (signal locked to the rubidium atomic standard) and issues a signal locked to it.
(1) Calibration specifications
• Reference oscillator (standard)
– Frequency
10 MHz
– Aging rate
±5 × 10
-9
/day
– Activation characteristics
× 10
-7
/10 min. (reference after 24 hour operation)
– Temperature characteristics
±3 × 10
-8
 (0 to 50
°C)
• Reference oscillator (Opt 01)
– Frequency
10 MHz
– Aging rate
× 10
-10
/day
– Temperature characteristics
±5 × 10
-9
 (0 to 50
°C)
(2) Measuring instruments for calibration
• Oscilloscope
• Frequency standard