Anritsu MG3641A User Manual

Page of 212
SECTION 4   OPERATING INSTRUCTIONS
4-50
4.14 Error Messages
If a panel operation or device internal error occurs, the MG3641A/MG3642A displays messages in the error message
area on the multimenu display.
Error messages are given below:
Operation error:  Displayed when a setting becomes invalid because of an incorrect operation.
Data out of range:
An attempt is made to set a value out of range
(e.g., to set the frequency to 10 GHz).
Invalid suffix:
A unit key not appropriate for a set parameter is pressed
(e.g., an attempt is made to set the sweep time to 1 MHz)
Execution error:
An attempt is made to execute one of the following incorrect operations:
•  To start executing the sweep with an incorrect sweep parameter specified.
•  To display too many digits on the display unit for the output level offset in
volts.
•  To exceed the upper limit of the variance range (
±10 dB) in level continuous
mode.
Not Stored memory:
An attempt is made to recall a memory not stored.
Trigger entry overflow:
An attempt is made to register a trigger program of more than 20 steps.
Trigger entry ignored:
An attempt is made to register the trigger program when sweep is being executed
or suspended.
Hardware error:
An attempt is made to execute settings relating to an unmounted option
(e.g., to turn on pulse modulation when the pulse modulator is not mounted.)
Main Menu (1/2)
Mod
Ofs
Mem
Data out of range
Swp
AF
Error messages