The  Counters  are  incremented  any  time  the  calibration  section  or  the  Factory 
parameters section have been modified.  
 
At power on, the display will show the current software revision number  followed 
by the message of the Calibration Count   “[AL[nt” then a number i.e.  “123”.   The 
number  from  the  counter  memory.      Then  the  Parameter  Counter  message  of 
“PAr[nt”    and  probably  a  different  number,  i.e.    “234”.      The  counters  cannot  be 
reset to 0, they will increment until the display can no longer hold the values.  (1 to 
999999).   It is expected we will never have more than 1 million calibrations in the 
life of the machine.    
Each display is held for 1-2 seconds. 
The scale will then continue to do the display test and go to normal weighing. 
If during the time the counting displays are shown, the user presses the [Tare] key, 
the  user  will  be  given  a  message  to  enter  the  passcode  necessary  to  calibrate  the 
scale, “P - - - - “   Enter the code  “P0000”  to Enter calibration or “P1000” to enter 
the parameters, followed by pressing the [Tare] key.   
The  Calibration  access  will  allow  user  calibration  (See  section  15.1)  and  the 
parameter code will allow access to the following parameters. (see section 15.2). 
 “F4 Int”
 
 
 
 
Initial Zero Range  
“F5 rEZ” 
 
 
 
Re-Zero range  
“F6 SCS” 
 
  
 
Successive Tare Enable 
“F7 Cnt” 
 
 
 
Display ADC counts 
“F8 Zem” 
 
 
 
Zero Mode   
“F9 Lvd” 
 
 
 
Low voltage detection