14.
 
ERROR MESSAGES 
During the initial power-on testing or during operation, the scale may show an error 
message.  The meaning of the error messages is described below. 
If an error message is shown, repeat the step that caused the message. If the error 
message is still shown then contact your dealer for support.
 
ERROR 
CODE
 
DESCRIPTION 
POSSIBLE CAUSES 
 
Err  1 
Time input Error 
Invalid  time  entry  such  as  “268970” 
for the time format “H-m-S”. 
Err  2 
Date input Error 
34th  day  of  a  month  is  an  invalid 
entry. 
Err  4 
Initial  Zero  is  greater  than 
allowed  (4%  of  maximum 
capacity)  when  power  is 
turned  on  or  when  the 
[Zero/Enter] key is pressed. 
Weight on the platform when turning 
the scale on. 
Excessive  weight  on  the  platform 
when zeroing the scale. 
Platform is not installed. 
Improper calibration of the scale. 
Damaged load cell. 
Damaged Electronics. 
Err 6 
A/D  count  is  not  correct 
when turning the scale on. 
Load cell is damaged. 
Electronics is damaged. 
Err 7 
Percent input error 
Percent  function  is  entered  with  no 
reference mass on the platform. 
Err  8 
High limit input error 
Low  limit  is  set  first,  then  the  high 
limit  is  set  lower  than  the  low  limit 
and high limit not equal to zero.