F4 –INITIAL ZERO RANGE 
To enter this parameter, press the [Tare] key when “F4 int” is shown. 
The display will show the current initial zero range.      
Press the [Func] key to change the value and press [Tare] to accept the value. 
Press [Zero] to return to weighing. 
 
F5 -RE-ZERO RANGE 
To enter this parameter, press the [Tare] key when “F5 rEZ” is shown. 
The display will show the current re-zero range.      
Press the [Func] key to change the value. 
Press [Tare] to accept the value. 
Press [Zero] to return to weighing. 
 
F6 -SUCCESSIVE TARE  
To enter this parameter, press the [Tare] key when “F6 SCS” is shown. 
The display will show if the successive tare is on or off. 
Press the [Func] key to change the value. 
Press [Tare] to accept the displayed value. 
Press [Zero] to return to weighing. 
 
F7 –ADC COUNTS  
To enter this parameter, press the [Tare] key when “F7 Cnt” is shown. 
This parameter allows you to view the A/D counts from the internal A/D converter. 
This can be an aid to service. 
Press the [Tare] key to return to the PARAMETER menu. 
Press the [Zero] key to return to weighing. 
Typical value at zero is 30,000-90,000 (approx.)  