Adam Equipment azplus 15 User Manual

Page of 32
© Adam Equipment Company 2007 
18
 
6.2.1 
F1 -CALIBRATION 
 
To enter the calibration parameter, press the [Tare] key when “
F1
” is displayed. 
 
The display will instruct you to remove any weight from the scale, “
UnloAd
”  
   
The display will tell you to add weight to the scale: “
LoAd
” “    
15
” “
kilo
” 
 
Add the weight shown, wait for stability then press the [Tare] key. 
 
The display will show “
SPAn
” “
PASS
” if the calibration is OK. Or it will show “
SPAn
” “
FAIL
” if 
calibration could not be completed correctly. 
 
The scale will return to the parameter menu. 
 
6.2.2 
F2 -RESOLUTION 
 
To enter this parameter, press the [Tare] key when “
F2
” is shown. 
 
The display will show the current resolution, either 3000 or 6000 or Dual Range.   
 
Press the [PLU4] key to change the value. 
 
Press [Tare] to accept the displayed value. 
 
NOTE: Only 3000 is allowed for the AZ
plus scales. 
 
6.2.3 
F3 -CAPACITY 
 
To enter this parameter, press the [Tare] key when “
F3
” is shown. 
 
The display will show the current capacity in 6, 15 or 30 kilograms.  
 
Press the [PLU4] key to change the value. 
 
Press [Tare] to accept the displayed value.