 Platform Scale 9K0U    Operator Guide
NOTE: The scale must be installed according to the instal-
lation instructions accompanying the scale before per-
forming the calibration procedure here. It is recommended 
that you check accuracy of the scale every three months 
with a known weight (for example, 5 lb.). The weight 
displayed should be the weight applied minus the mainte-
nance tolerance.
To check accuracy (Weight Classifier mode; default for 
USPS) perform one of the following tests:
• 
Place a 5 lb. Test Weight on scale platter. The weight 
should read 4 lb. 15.80 oz.  If the weight reads less 
than 4 lb. 15.6 oz. or greater than 5 lb. 0.00 oz. then the 
scale is out of tolerance and must be recalibrated.
• 
Place a 10 lb. Test Weight on scale platter. The weight 
should read 9 lb. 15.8 oz.  If the weight reads less than 
9 lb. 15.4 oz. or greater than 10 lb. 0.00 oz. then the 
scale is out of tolerance and must be recalibrated.
• 
Place a 20 lb. test weight on scale platter. The weight 
should read 19 lb. 15.6 oz.  If the weight reads less 
than 19 lb. 15.2 oz. or greater than 20 lb. 0.00 oz. then 
the scale is out of tolerance and must be recalibrated.
It is recommended that calibration be performed once a 
year.
Contact your local District to obtain a 20 lb. calibration weight 
or contact your Pitney Bowes Customer Service Representa-
tive to have the scale calibrated.
To   calibrate your scale:
1.  The scale should be level and powered up a minimum of 
ten (10) minutes prior to calibration.
2. Press the Setup key. The scale displays the user setup 
menu.
3. Select Service Setup from the menu. At the Enter Pass-
word prompt, enter the following key sequence:
 
SERREP
Calibrating the 
9K0U Scale 
(20 lb. Calibration 
Weight Required)