ERROR MESSAGES
The apparatus registers any problems and shows the following error messages
on the display:
30
Error message
Meaning
Cause
Solution
ER-1
Error during 
1. The armband is not correctly attached.
1. Attach the armband correctly.
pumping
2. The armband is broken.
2. Contact the retailer.
3. The pump or valves are defective.
3. Contact the retailer.
ER-2
Disturbances or  
1. The user moves or speaks
1. The user must sit still and avoid talking during the reading.
noise during readings during the reading.
2. The user is too nervous.
2. Have a blood pressure 
reading taken by a doctor.
ER-3
Error while emptying  1. The valves are defective.
1. Contact the retailer.
the armband
2. The armband is not correctly attached.
2. Attach the armband correctly.
ER-4
Abnormal reading
The user may have a heart problem.
Consult a doctor.
HI
The pressure in the  
The air tube is blocked, the pump 
Try to take another reading, 
armband is too high.
mechanism is defective or the pressure 
or contact the retailer.
monitor is defective.
Battery symbol
Flat batteries
The batteries are almost dead.
Replace the batteries.