MEDIUM DISTANCE, INDOOR/OUTDOOR
INSTALLATION
When an outdoor installation is required, it is necessary to install a
transmitter to obtain accurate readings at distances from 10 to 50 m
(33-165').
Since the introduction of AmpHel
®
 these distances are no longer a
problem. You can now  connect your meter directly to an AmpHel
®
electrode, saving the cost of a transmitter.
The standard cable length of the AmpHel
®
 electrode is 5 m (16.5').
Additional lengths of regular cable up to 50 m (165'), can be
installed without special connectors. Coaxial cables are always pre-
ferred over normal cables due to their excellent insulation, even
though, Amphel
®
 electrodes can operate with both.
AmpHel
®
 electrodes have a micro-amplifier in the electrode cap to
boost the signal, drastically reducing susceptibility to noise and drift.
With all of the components sealed in the electrode body, moisture up
to 100% RH will not effect the signal.
pH METER
WITH AN AmpHel
ELECTRODE
MEASUREMENTS CAN BE
TAKEN FROM DISTANCES UP TO
50 METERS (165 FEET)
WET ENVIRONMENT
RELATIVE HUMIDITY UP TO 100%
MAXIMUM 50 METERS
®
pH-ORP METER
pH CALIBRATION SOLUTIONS
HI 7004M
pH 4.01 buffer solution, 230 mL
HI 7004L
pH 4.01 buffer solution, 460 mL
HI 7006M
pH 6.86 buffer solution, 230 mL
HI 7006L
pH 6.86 buffer solution, 460 mL
HI 7007M
pH 7.01 buffer solution, 230 mL
HI 7007L
pH 7.01 buffer solution, 460 mL
HI 7009M
pH 9.18 buffer solution, 230 mL
HI 7009L
pH 9.18 buffer solution, 460 mL
HI 7010M
pH 10.01 buffer solution, 230mL
HI 7010L
pH 10.01 buffer solution, 460 mL
ORP SOLUTIONS
HI 7020M
200-275mV ORP solution, 230 mL
HI 7020L
200-275mV ORP solution, 460 mL
HI 7091M
Pretreatment reducing sol., 230 mL
HI 7091L
Pretreatment reducing sol., 460 mL
HI 7092M
Pretreatment oxidizing sol., 230 mL
HI 7092L
Pretreatment oxidizing sol., 460 mL
ACCESSORIES
ACCESSORIES
ACCESSORIES
ACCESSORIES
ACCESSORIES
The probe can be compensated for normal contamination by a process
of re-calibration. However, it is recommended that the process conduc-
tivity probe be removed from the system regularly for maintenance.
Rinse the HI 7637     probe with tap water. If a more thorough
cleaning is desired, unscrew the PVC sleeve and clean the sensors with
a non-abrasive cloth and alcohol. When reinstalling the PVC sleeve
onto the probe, be sure that the holes are towards the top of the
probe (the end with the cable).
After cleaning, re-calibrate the instrument. However, if the instrument
will not re-calibrate, replace the probe.
Note: Always re-calibrate the meter after the replacement of the
probe.