Hanna Instruments wm 8910 User Manual

Page of 29
1 1
1 0
10. Output contacts:
– Two terminals marked "°C Sensor" for the connection of a
temperature probe for automatic temperature compensa-
tion
– Two terminals marked "mA Output" for the isolated output
BOTTOM CONNECTIONS
1. BNC socket for a combination pH electrode
2. Input voltage supply socket
3. Output voltage socket for acid dosing (supplies power to a
pump or electrovalve). This socket is activated by a relay
when the pH is higher than the "SET HI" value.
4. Output voltage socket for alkaline dosing (supplies power to a
pump or electrovalve). This socket is activated by a relay
when the pH value is lower than the "SET LO" value.
Unplug the instrument from the power supply before
replacing the fuse.
Specifications
WM 8911D
                 WM 8911U
RANGE
0.00 to 14.00 pH
RESOLUTION
0.01 pH
ACCURACY
±0.02 pH
(@20°C/68°F)
TYPICAL EMC
±0.1 pH
DEVIATION
mA OUTPUT
User selectable 0 to 20 mA or 4 to 20 mA
in a 0-14 pH range with isolated output
OFFSET
Through "OFFSET' trimmer
CALIBRATION
(Max. ±1.5 pH regulation)
SLOPE
Through "SLOPE" trimmer
CALIBRATION
from 80% to 110%
TEMPERATURE
Manual or automatic
COMPENSATION
from -10 to 80°C (14 to 176°F)
SETPOINT RANGE
From 0.00 to 14.00 pH with 2 trimmers:
"COARSE" for an approximate
regulation, "FINE" for fine tuning.
ALARM CONTACT
Through a relay which provides a normally open
 or a normally closed contact output (isolated
output Max.2A, Max.240V, resistive load, 1,000,000
 strokes). The alarm contact will be activated if the
pH varies by -2.0 pH from "SET LO" and
+2.0 pH from "SET HI" points.
POWER SUPPLY
220/240V
110/115V
50/60Hz
50/60Hz
POWER SUPPLIES
One plug for acid and another for alkaline dosage
FOR  DOSING
(115 or 220V, depending on the model,
Max.2A, 1,000,000 strokes)
ENVIRONMENT
-10 to 50°C (14 to 122°F)
max. 85% RH non-condensing
PROTECTION
IP54
WEIGHT
1.6 Kg (3.5 lb.)
ENCLOSURE
181 x 221 x 142mm (7.1 x 8.7 x 5.6")
MATERIAL
Housed in rugged, fiber-reinforced polypropylene case