Omega OMB-DAQBOARD-3000 User Manual

Page of 108
 
Using the Hysterisis Function 
Update Mode: N/A, the Hysterisis option has a forced update built into the function 
Criteria Used: window criteria for above and below the set limits 
 
 
The figure below shows analog input Channel 3 with a setpoint which defines two 16-bit limits, Limit A 
(High) and Limit B (Low).  These are being applied in the hysteresis mode and DAC Channel 0 will be 
accordingly.   
In this example Channel 3’s analog input voltage is being used to update DAC0 as follows: 
o
  When outside the window, low (below Limit B) DAC0 is updated with 3.0V.  This 
update will remain in effect until the analog input voltage goes above Limit A. 
o
  When outside the window, high (above Limit A) DAC0 is updated with 7.0V.  This update will 
remain in effect until the analog input signal falls below Limit B.  At that time we are again 
outside the limit “low” and the update process repeats itself.  
 
Hysteresis mode can also be done with P2C digital output port, or a timer output, instead of a DAC. 
 
 
 
 
Limit A 
Limit B 
 
Detection 
DAC0 
 
 
Channel 3 in Hysterisis Mode  
 
Ch 3 Analog 
Input Voltage 
 
 
DaqBoard/3000 Series User’s Manual
 908794
 
Setpoint Configuration for Output Control      6-9