Hanna Instruments hi 983301 User Manual

Page of 2
Instruction Manual
  &  
HI 983302N
HI 983301N
HI 983301N/5
On-line, Waterproof
EC-TDS meter with large  LCD
http://www.hannainst.com
These Instruments are in
Compliance with the CE Directives
Hanna Instruments reserves the right to modify the design, construction
and appearance of its products without advance notice.
WARRANTY
EC and TDS  
  are warranted for one year against
defects in workmanship and materials when used for their intended
purpose and maintained according to instructions. The probes are
warranted for a period of six months. This warranty is limited to
repair or replacement free of charge.
Damages due to accident, misuse, tampering or lack of prescribed
maintenance are not covered.
If service is required, contact the dealer from whom you purchased the
instrument. If under warranty, report the model number, date of pur-
chase, serial number and the nature of the failure. If the repair is not
covered by the warranty, you will be notified of the charges incurred. If
the instrument is to be returned to Hanna Instruments, first obtain a
Returned Goods Authorization Number from the Customer Service de-
partment and then send it with shipment costs prepaid. When shipping
any instrument, make sure it is properly packaged for complete protec-
tion.
To validate your warranty, fill out and return the enclosed warranty card
within 14 days from the date of purchase.
All rights are reserved. Reproduction in whole or in part is prohibited
without the written consent of the copyright owner, Hanna Instru-
ments Inc., Woonsocket, Rhode Island, 02895 , USA.
Dear Customer,
Thank you for choosing a Hanna product. This manual will
provide you with the necessary information for a correct
operation. Please read it carefully before using the meter. If
you need additional technical information, do not hesitate
to e-mail us at tech@hannainst.com.
These instruments are in compliance with the 
 directives.
Remove the instrument from the packing material and
examine it carefully. If any damage has occurred during
shipment, immediately notify your Dealer or the nearest
Hanna Customer Service Center.
The meters are supplied with:
• HI 7632 EC probe (HI 983302N );
• HI 7634 TDS probe (HI 983301N and HI 983301N/5);
• 1413 µS/cm calibration solution (20 mL) for HI 983302N;
• 1500 ppm calibration solution (20 mL) for HI 983301N;
• 1382 ppm calibration solution (20 mL) for HI 983301N/5;
• Calibration screwdriver;
• 12 VDC power adapter.
Note: Conserve all packing material until the instrument
has been observed to function correctly. Any defective
item must be returned in its original packing.
PRELIMINARY EXAMINATION
EC  and TDS 
  are instruments specially de-
signed to meet the needs of growers in greenhouses hydroponic
applications. They are equipped with a large Liquid Crystal
Display for an easy reading also from distance.
You can simply hang the meter right above the sample to
be tested for continuous measurement and the unit will run
without interruption on 12 VDC power supply.
The housing has been completely sealed against vapors and
humidity.
The meters are supplied with a probe that automatically
compensates for the temperature variation. The probe is
easy to clean and requires little maintenance.
Measurements are highly accurate and the meters can be
calibrated at one point.
In addition, the two TDS 
  use a specially
devised formula which perfectly reflects the needs of growers.
In fact, the common method of measuring Total Dissolved
Solids by multiplying the conductivity by a converting value
is far off the mark in horticulture with high concentrations of
phosphate, nitrogen, etc.
In the HI 983301N model the TDS factor is 0.7 while in the
HI 983301N/5 is 0.5.
GENERAL DESCRIPTION
FUNCTIONAL DESCRIPTION
1. Molded eye
2. Calibration trimmer
3. EC or TDS probe
4. 12 VDC power adapter
5. Power supply connector
6. Liquid Crystal Display (LCD)
SPECIFICATIONS
EC 
 – HI 983302N
Range
0.00 to 9.99 mS/cm (EC)
Resolution
0.01 mS/cm (EC)
Accuracy (@ 25°C/77°F)
±2% f.s.
Typical EMC Deviation
±2% f.s.
Temp.Compensation
Automatic
5 to 50°C (41 to 122°F)
Calibration
Manual with one trimmer
Probe
HI 7632 EC probe (included)
Casing
IP54
Power supply
External 12 VDC (included)
Dimensions
86 x 94 x 33 mm (3.4 x 3.7 x 1.3”)
Weight
150 g (5.3 oz)
TDS 
 – HI 983301N and HI 983301N/5
Range
0 to 1990 mg/L (ppm)
Resolution
10 mg/L (ppm)
Accuracy (@ 25°C/77°F)
±2% f.s.
Typical EMC Deviation
±2% f.s.
TDS Factor
HI 983301N
0.7
HI 983301N/5 0.5
Temp.Compensation
Automatic
5 to 50°C (41 to 122°F)
Calibration
Manual with one trimmer
Probe
HI 7634 TDS probe (included)
Casing
IP54
Power supply
External 12 VDC (included)
Dimensions
86 x 94 x 33 mm (3.4 x 3.7 x 1.3”)
Weight