Hanna Instruments bl 931700 User Manual

Page of 2
WARRANTY
Instruction Manual
BL 931700-0
BL 931700-1
Panel-Mounted
pH Indicators &
Controllers
These instruments are warranted for two years against defects in work-
manship and materials when used for their intended purpose and
maintained according to instructions. Probes are warranted for six months.
This warranty is limited to repair or replacement free of charge.
Damages due to accident, misuse, tampering or lack of prescribed main-
tenance are not covered. If service is required, contact the dealer from
whom you purchased the instrument. If under warranty, report the model
number, date of purchase, serial number and the nature of the failure. If
the repair is not covered by the warranty, you will be notified of the
charges incurred. If the instrument is to be returned to Hanna Instruments,
first obtain a Returned Goods Authorization Number from the Customer
Service department and then send it with shipment costs prepaid. When
shipping any instrument, make sure it is properly packaged for complete
protection.
BL931700-0 and BL931700-1 are panel-mounted pH indi-
cators and controllers designed for simplicity of use in a wide
range of industrial applications.
Connections and wiring to electrode, power supply, contacts
and external recorders are made via the terminal blocks on
the rear panel. The meters are equipped with a BNC socket
and accept input from conventional pH electrodes.
Other features include: 4-20 mA output, overtime control
system, selection of dosing direction (Acid/Alk), contact for
external control and disabling of dosing action, multi-colour
LED for indicating if the meter is in measurement/dosing/
alarm condition, possibility to set (Off-Auto-On switch) dosing
action mode.
Two models are available:
• BL 931700-0 powered at 12 Vdc
• BL 931700-1 powered at 115 or 230 Vac
GENERAL DESCRIPTION
SPECIFICATIONS
Range
0.00 to 14.00 pH
Resolution
0.01 pH
Accuracy 
(@ 20°C/68°F)
±0.02 pH
Typical EMC Deviation
±0.10 pH / ±0.10 mA
Calibration
 Manual, through Offset & Slope trimmers
Dosing Contact
Maximum 2A (fuse protected),
250 Vac, 30 Vdc
Dosing Selection
Acid or Alk, selectable on the back panel
Contact Open=Acid dosage=Relay ON if Measure > Setpoint
Contact Close=Alk dosage=Relay ON if  Measure < Setpoint
Setpoint
Adjustable, from 0 to 14 pH
Overtime
Adjustable, typically from 5 to approx. 30 minutes
Recorder Output
4 to 20 mA, accuracy ±0.20 mA,
500
 Ω
 maximum load
Power Consumption
10 VA
Installation Category
II
Power supply:
External (fuse protected)
BL931700-0
12 Vdc
BL931700-1
115/230 Vac ; 50/60Hz
Dimensions
83 x 53 x 99 mm (3.3x2.1x3.9”)
ACCESSORIES
pH CALIBRATION SOLUTIONS
HI 7004M/L pH 4.01 buffer solution, 230/500 mL bottle
HI 7006M/L pH 6.86 buffer solution, 230/500 mL bottle
HI 7007M/L pH 7.01 buffer solution, 230/500 mL bottle
HI 7009M/L pH 9.18 buffer solution, 230/500 mL bottle
HI 7010M/L pH 10.01 buffer solution, 230/500 mL bottle
OTHER SOLUTIONS
HI 70300M/L Electrode storage solution, 230/500 mL bottle
HI 7061M/L Electrode general cleaning solution, 230/500 mL
HI 7073M/L Protein cleaning solution, 230/500 mL bottle
HI 7074M/L Inorganic cleaning solution, 230/500 mL bottle
HI 7077M/L Oil & Fat cleaning solution, 230/500 mL
REFILLING ELECTROLYTE SOLUTIONS (50 mL, 4 pcs.)
HI 7071
3.5M KCl+AgCl, for single junction electrodes
HI 7072
1M KNO
3
 solution
HI 7082
3.5M KCl, for double junction electrodes
pH ELECTRODES
HI 1002/5
BNC connector, double junction, plastic-body
w/external thread & 5 m (16.5’) cable
HI 1090T
Screw connector, double junction, glass-body
with external thread
HI 1110S
Screw connector, single junction, glass-body
HI 1210T
Screw connector, double junction, plastic-body
with external thread
HI 2114P/2 BNC connector, double junction, plastic-body
with 2 m (6.6’) cable
HI 1210B/5 BNC connector, double  junction, plastic-body,
with 5 m (16.5') cable
HI 2910B/5 BNC connector,  double  junction, plastic-body,
with built-in amplifier and  5 m (16.5') cable
EXTENSION CABLES, SCREW TO BNC CONNECTOR
HI 7855/5
Extension cable 5 m (16.5') long
HI 7855/10 Extension cable 10 m (33') long
OTHER ACCESSORIES
BL PUMPS Dosing Pumps (flow rate from 1.5 to 20LPH)
HI 6050
Submersible  electrode holder, 60 cm (24”)
HI 6051
Submersible electrode holder, 110 cm (43”)
HI 6054B
Electrode holder for in-line applications
HI 6054T
Electrode holder for in-line applications
HI 710005/6 12 Vdc power adapter, US/European plug
HI 731326
Calibration screwdriver (20 pcs)
HI 740146     Mounting brackets
HI 7871
Level Controller (min and max)
HI 7873
Level Controller (min, max and overflow)
Remove the instrument from the packing material and
examine it carefully. If any damage has occurred during
shipment, immediately notify your Dealer or the nearest
Hanna Customer Service Center.
Each meter is supplied with:
• Mounting brackets
• Instruction manual
Note: Conserve all packing material until the instrument
has been observed to function correctly. Any defective
item must be returned in its original packing.
Dear Customer,
Thank you for choosing a Hanna product.
This manual will provide you with the necessary information
for the correct operation of the meter. Please read it carefully
before using the instrument.
If you need additional technical information, do not hesitate
to e-mail us at tech@hannainst.com.
These instruments are in compliance with the 
 directives.
PRELIMINARY EXAMINATION
ASSEMBLING VIEW
Recommendations for Users
Before using these products, make sure that they are entirely suitable for the environment
in which they are used. Operation of these instruments in residential areas could cause
unacceptable interferences to radio and TV equipment. The glass bulb at the end of the
electrode is sensitive to electrostatic discharges. Avoid touching this glass bulb at all times.
During operation, ESD wrist straps should be worn to avoid possible damage to the electrode
by electrostatic discharges. Any variation introduced by the user to the supplied equipment
may degrade the instrument’s EMC performance. To avoid electrical shock, do not use these
instruments when voltages at the measurement surface exceed 24 Vac or 60 Vdc. To avoid
damages or burns, do not perform any measurement in microwave ovens.